# Multi-Parameter Patient Monitor

Category: ICU/CCU

Price: Call or Whatsup us for price

Product Short Description:

Multi-Parameter Patient  Monitor 

Model: PM-7000D
Brand: Zoncare
Origin: China


● Original perfusion index (Pl) and show signal quality of blood oxygen infusion.

● Activated ETCO2 port, optional ETCO2 module plug and play on different patient monitors One key reset alarm preset values.

● Arrhythmia analysis, ST segment analysis and drug concentration and titration table calculation, ST base dot automatically adjustment.

● Defibrillation and resistance electrome design.


Product Description:

ECG

Meet the standards

ANSI/AAMI EC13:2002, IEC 60601-2-27:2011

Lead configuration

Standard three-guide or five-guide cable:

Three-guide: RA、LA、LL, Method of leads: I,II,III;
Five-guide: RA、LA、LL、RL、V,

Method of leads: I,II,III,aVR,aVL,aVF,V

Lead standard

AHA、IEC

display sensitivity

At least 1.25 mm/mV (×0.125), 2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/mV (×1), 20 mm/mV (×2) and automatic gain are supported, with an error of less than ±5%.
Add ±300 mV DC polarization voltage, sensitivity variation range ±5%

Wave scanning speed

At least three scanning speeds of 50mm /s, 25mm /s and 12.5mm /s can be provided with an error of no more than ±10%.


RESP

Measurement mode

Thoracic Electrical Bioimpedance

Measure the lead

Lead I and II are optional. Default is lead II

Bandwidth

0.2~2.5 Hz(-3dB)

Scanning speed

12.5 mm/s、25mm/s、50 mm/s

RR range

Adults:0 rpm~120 rpm

Children、newborn:0 rpm~150 rpm

RR resolution

1 rpm

Accuracy

8 RPM ~ 150 RPM: ±2 RPM or ± 2%, whichever is larger
Other scope is not defined.

Suffocation alarm delay

10 s、15 s、20 s、25 s、30 s、35 s、40 s


SpO2

Oxygen saturation measurement

Measurement range: 0% ~ 100%; Resolution: 1%

Accuracy

70% ~ 100%: error ≤±2%;
Within the range of 0% to 69%: not defined.

Replacement cycle

1 s

Perfusion index

0.05%~ 20%, precision undefined

Blood oxygen saturation alarm setting range

1) Alarm high limit: 2% ~ 100%;
2) Low alarm limit: 0% ~ 98%;
3) Alarm error: ±1%.


NIBP

Meet the standard

IEC 60601-2-30, IEC 60601-2-49,ANSI/AAMI SP10

Measurement mode

succussion

Operating mode

Manual/automatic/continuous measurement

Automatic measurement mode measurement interval time

1,2,3,4,5,10,15,30,60,90,120,180,240,480 minutes

Continuous mode measurement time

5 min


PR

Measurement range

30bpm~250bpm

Resolution

1bpm

Accuracy

±2bpm or 2%

Alarm limit

1)Adult

Alarm high limit: 17bpm~300bpm;

Alarm low limit: 15bpm~295bpm;
2) Pediatric

Alarm high limit: 17bpm~350bpm;

Alarm low limit: 15bpm~345bpm;

3) Neonate

Alarm high limit: 17bpm~350bpm;

Alarm low limit: 15bpm~345bpm;